WASHINGTON, DC — The Food and Drug Administration says Tandem Diabetes is recalling certain Tandem Mobi insulin pumps because some devices may produce false occlusion alarms even when insulin is flowing normally. In those cases, the pump can suspend insulin delivery until the user acknowledges the alarm and restarts therapy.
The agency classifies the action as the most serious type of recall. FDA officials say continuing to use an affected pump without following the company’s instructions could lead to serious injury or death.
Tandem has already sent letters to customers, health care providers, distributors, and pharmacies explaining which devices need replacement and which can remain in use with updated instructions.
Which Tandem Mobi pumps are included in the recall
The recall covers all Tandem Mobi insulin pumps with interoperable technology, but not every device is being handled the same way. Tandem says some pumps are on a serial-number list for replacement, while others remain in service if users follow new safety steps.
Customers whose pumps appear on the replacement list were told to confirm the device is affected and immediately request a return and replacement through the link in the company letter or by calling Tandem Customer Technical Support at 877-801-6901.
Users whose serial numbers are not on that list may continue using the pump, but they are being told to follow the revised instructions closely and watch for repeated alarm issues.
What users with replacement pumps are being told to do
Tandem’s Sept. 23 letters instructed affected customers to stop relying on the targeted pumps and begin the replacement process right away. The company also told them to pay close attention to occlusion alarms and to inspect the cartridge, tubing, and infusion site whenever an alarm occurs.
Users are supposed to acknowledge the alarm, correct any blockage or damage found, and then resume insulin delivery. Tandem also says people should enable push notifications on the Tandem Mobi mobile app so they can receive an alert if an alarm sounds.
The pump signals an occlusion alarm through audible tones or vibrations, depending on settings, and the warnings grow louder until the alarm is acknowledged.
Instructions for other Tandem Mobi users still allowed to keep pumping
For all other Tandem Mobi users, the company says the pump may continue to be used while the updated instructions are followed. Those instructions mirror the same alarm response steps: acknowledge the warning, inspect the system, fix any problem found, and restart insulin delivery.
Tandem says users should also keep smartphone notifications turned on in the mobile app so alarm alerts are not missed. If repeated occlusion alarms keep happening after troubleshooting, customers are supposed to contact Tandem Customer Support.
The company also warns all users to monitor blood sugar regularly and keep a backup insulin delivery method ready, as directed by their health care provider and pump training materials.
Why the FDA says the recall matters for diabetes management
Tandem says the problem can make the pump incorrectly believe there is an insulin flow obstruction. When that happens, insulin delivery stops until the alarm is cleared and therapy is resumed, creating the risk of repeated interruptions.
Those interruptions can lead to elevated blood glucose levels. In severe cases, the company says hyperglycemia may require medical treatment or hospitalization.
The Tandem Mobi pump is designed to deliver insulin under the skin for diabetes management and can work with compatible continuous glucose monitors and automated insulin dosing systems to adjust insulin delivery.
Reports of serious injury and how patients can get help
As of Oct. 28, 2026, Tandem Diabetes had reported 1,248 serious injuries connected to the issue and no deaths. The FDA says customers in the United States can report adverse reactions or quality problems through the agency’s MedWatch program.
People with questions about the recall, adverse reactions, or product quality concerns can contact Tandem Diabetes at [email protected] or 877-801-6901.
The FDA is also posting company-provided information, including replacement guidance and a support article about turning on smartphone notifications for the mobile app. The agency says the unique device identifier system helps track individual medical devices more quickly when problems arise.